ETHICON ENDO-SURGERY, LLC. LIGACLIP* APPLIER W/MECH STOP; APPLIER, SURGICAL, CLIP Back to Search Results
Catalog Number LX107
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a thyroidectomy surgery.After clamped the clips and clip could not close.Changed another one to complete the procedure.There was no patient consequence reported, no additional information could be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent 6/10/2025.D4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4) date sent 7/15/2025.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the lx107 device was received with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon functional testing of the device, the instrument loaded, retained, and deployed 6 clips as intended.The instrument was fully functional and conforming.The event described could not be confirmed as the device performed without any difficulties noted.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: select the appropriate size ligaclip extra ligating clip cartridge and corresponding clip applier.With the cartridge slots facing away from the base, insert the cartridge into one of the channel openings of the lc800 stainless steel cartridge base.Ensure that the cartridge is past the channel opening and securely held in place.Grasp the applier in the box lock area using the pencil grip technique.Insert the jaws of the instrument into the individual cartridge slot, making sure the tips of the applier are perpendicular to the surface of the cartridge.Insert the applier until it stops.Do not force the applier.It should enter and withdraw from the cartridge smoothly.Keeping the applier jaws perpendicular to the surface of the cartridge, retract the applier from the cartridge.The clip will be securely held in the applier jaws.It is not necessary to maintain ring tension to hold the clip in the applier jaws.Position the clip around the tubular structure to be ligated.Apply sufficient force to fully close the applier to assure that the clip is satisfactorily placed and secured.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Sn: (b)(6).An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 7/9/2025.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
3035526892
MDR Report Key22178980
Report Number3005075853-2025-04401
Device Sequence Number8906748
Product Code GDO
UDI-Device Identifier20705036012853
UDI-Public(01)20705036012853(17)401231(10)FQGA13062
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberLX107
Device Lot NumberFQGA13062
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/06/2025
Supplement Date Received by Manufacturer07/08/2025
07/08/2025
Initial Report FDA Received Date06/10/2025
Supplement Report FDA Received Date07/09/2025
07/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-