| Catalog Number |
G310SCLR |
| Medical Device Problem Code |
Restricted Flow rate (1248)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
While using a cavitron 300 series g310, they allege that they have intermittent water and vibration and the handpiece is heating up; no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
Hpc# 09190, sterimate# 202406.W4d water sol,iso valve build up/debris the water solenoid fails to maintain the preset pressure due to debris buildup.Soiled water tubing inside the unit affecting water quality.Hardened and deteriorated water supply hose with debris buildup in the water filter will replace damaged /worn components and recalibrate unit to factory specs upon estimate approval.Note: loaner included in the estimate.
|
| |
|