• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION INC. SOFIA FLOW 5F; CATHETER INTRAVASCULAR DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION INC. SOFIA FLOW 5F; CATHETER INTRAVASCULAR DIAGNOSTIC Back to Search Results
Model Number DA6125ST-ASP-US-OD
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intracranial Hemorrhage (1891)
Date of Event 03/21/2024
Type of Reportable Event Serious Injury
Event or Problem Description
As reported through a restore clinical study, the patient had subarachnoid hemorrhage post the mechanical thrombectomy procedure using an aspiration catheter.The patient experienced a subarachnoid hemorrhage noted on ct scan and was described as mild, and asymptomatic.There was no intervention performed.Spontaneously, the event outcome resolved without sequelae by (b)(6)2024.The relationship to the device is considered possibly.The relationship to the mechanical thrombectomy procedure is considered definitely.The relationship to the study disease is considered possibly.The relationship to the concurrent condition/treatment is considered possibly.
 
Additional Manufacturer Narrative
Investigation findings: the physical device was not available for evaluation to investigate if a condition existed that would have contributed to the reported event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation further examine the cause of the reported event.This information may be updated if additional information is provided at a later date.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Based on a review of the device¿s risk documentation, the reported event did not indicate there were the presence of any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform an analysis of the device, this investigation cannot identify with certainty any potential root causes.Ifu review (additional information can be found in the ifu): potential complications.Potential complications include but are not limited to: vasospasm, hematoma at the site of entry, embolism, ischemia, intracerebral/intracranial hemorrhage, pseudo aneurysm, seizure, stroke, infection, vessel dissection, thrombus formation, and death.Precautions.Exercise care in handling the sofia flow plus aspiration catheter to reduce the chance of accidental damage.Use caution when manipulating the sofia flow plus aspiration catheter in tortuous vasculature to avoid damage.Avoid advancing or withdrawal against resistance until the cause of resistance is determined.The presence of calcification, irregularities, or other devices may damage the sofia flow plus aspiration catheter and potentially affect its insertion or removal.Maintain perfusion of heparinized saline for inner lumen of the sofia flow plus aspiration catheter to prevent thrombus formation.Do not use automated high-pressure contrast injection equipment with the sofia flow plus aspiration catheter because it may damage the catheter.The catheter has hydrophilic coating over the distal 60 cm.Make sure to hydrate the coating before use with heparinized saline.Once the catheter is hydrated, do not allow it to dry.Delivery of the sofia flow plus aspiration catheter 6.Navigation through the vasculature b.Insert the guidewire and / or microcatheter into the sofia flow plus aspiration catheter and advance the guidewire / microcatheter until the guidewire / microcatheter and the sofia flow plus aspiration catheter are aligned at the distal end.C.Using the introducer sheath provided in the package, carefully insert the sofia flow plus aspiration catheter and the guidewire through a hemostatic valve of the femoral sheath.Warning: do not over-tighten the hemostatic valve on the sheath or guide catheter through which the sofia flow plus aspiration catheter is inserted.Over-tightening may result in damage to the sofia flow plus aspiration catheter.E.Under fluoroscopic guidance, advance or withdraw the sofia flow plus aspiration catheter over the guidewire and / or microcatheter until the desired position is attained.Select vessels by slowly torqueing the sofia flow plus aspiration catheter if necessary.Warning: do not advance or withdraw the device when excessive resistance is met until the cause of resistance is determined.Warning: torqueing the sofia flow plus aspiration catheter excessively while kinked may damage the device resulting in separation of the device.Withdraw the entire device (the device, microcatheter, and guidewire) if the device is severely kinked.Aspiration through the sofia flow plus aspiration catheter 8.Under fluoroscopic guidance, position the distal tip of the sofia flow plus aspiration catheter at the desired vessel location.Warning: do not advance or withdraw the device when excessive resistance is met until the cause of resistance is determined.Warning: torqueing the sofia flow plus aspiration catheter excessively while kinked may damage the catheter resulting in separation of the catheter.Withdraw the catheter system, including the sofia flow plus aspiration catheter, microcatheter, and guidewire if the sofia flow plus aspiration catheter is severely kinked.9.Attach aspiration tubing to the aspiration pump and turn on the pump (refer to the directions for use of the aspiration tubing and aspiration pump manual).Confirm that the aspiration gauge reads -20 inhg.Ensure the stopcock on the aspiration tubing is in the closed position.13.To begin aspiration, turn the aspiration tubing stopcock to the open position and check to see if blood, thrombus, or embolus are aspirated through the system.14.If after 10 seconds blood is still observed flowing through the system, stop aspiration.To stop aspiration, turn aspiration tubing stopcock to the closed position.Carefully reposition the distal tip of the sofia flow plus aspiration catheter to engage the thrombus and resume aspiration.15.If flow is restricted or absent, maintain aspiration to make sure any thrombus or embolus is fully engaged with the distal tip of the sofia flow plus aspiration catheter.With the thrombus or embolus fully engaged, slowly pull back the sofia flow plus aspiration catheter and completely withdraw out of the patient.Warning: excessive aspiration with the distal tip of the sofia flow plus aspiration catheter engaged with vessel wall may cause vessel injury.16.With the sofia flow plus aspiration catheter outside of the patient body, flush the catheter to clear the catheter of any thromboembolic material that may be inside.Warning: do not attempt to clear inner lumen of the sofia flow plus aspiration catheter by infusion while keeping the device in the patient body.Remove the sofia flow plus aspiration catheter from the patient body before attempting to clear the lumen.17.If repeated access to the vasculature with the same device is desired, flush and clean the inner lumen of the device by infusion.Inspect the device for any damage.Reintroduce the sofia flow plus aspiration catheter into the body and follow steps 6 & 7 in the ¿delivery of the sofia flow plus aspiration catheter¿ section to navigate the catheter to the target site.Warning: do not use the device if any damage or irregularities are observed.This study is ongoing and device/procedure relatedness determinations can be re-adjudicated by independent reviewers/physicians.The study device and study procedure relatedness can change as a result of these reviews.Microvention is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by microvention, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Microvention has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, microvention, or its employees that the device, microvention, or its employees caused or contributed to the event described in the report.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFIA FLOW 5F
Common Device Name
CATHETER INTRAVASCULAR DIAGNOSTIC
Manufacturer (Section D)
MICROVENTION INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key22180362
Report Number2032493-2025-90243
Device Sequence Number21139116
Product Code DQO
UDI-Device Identifier00842429108475
UDI-Public(01)00842429108475(11)240109(17)261231(10)0000486575
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K150366
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDA6125ST-ASP-US-OD
Device Lot Number0000486575
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2025
Initial Report FDA Received Date06/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-