• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CUTERA, INC. TRUSCULPT ID; MASSAGER, VACUUM, RADIO FREQUENCY INDUCED HEAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CUTERA, INC. TRUSCULPT ID; MASSAGER, VACUUM, RADIO FREQUENCY INDUCED HEAT Back to Search Results
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Scar Tissue (2060)
Date of Event 06/11/2023
Type of Reportable Event Serious Injury
Event or Problem Description
An adverse event occurred in june 2023 involving burns to a patient's abdomen during a trusculpt treatment.An investigation was initiated at that time in collaboration with the treating clinic.The burns were reported by the clinic to be approximately 80-95% healed during follow-up and based on clinical assessment and available information at the time, the incident was not deemed to meet fda reportability criteria.The device was not returned or evaluated, as there was no indication of a malfunction or failure.The root cause of the incident was determined to be use error.Updated information: in may 2025, the clinic provided a follow-up photo of the treated area, which showed evidence of scar formation at the sites of the original burns.No new or ongoing adverse events have been reported by the clinic since the initial incident in (b)(6) 2023.Conclusion: the event has been reassessed in light of the updated clinical information, and this report is being submitted in accordance with fda requirements.The device functioned as intended, and no device-related malfunction contributed to the outcome.The root cause remains attributed to use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUSCULPT ID
Common Device Name
MASSAGER, VACUUM, RADIO FREQUENCY INDUCED HEAT
Manufacturer (Section D)
CUTERA, INC.
3240 bayshore blvd.
brisbane CA 94005
Manufacturer Contact
julia brown
3240 bayshore blvd.
brisbane, CA 94005
MDR Report Key22181031
Report Number2954354-2025-00005
Device Sequence Number8570448
Product Code GEI
UDI-Device Identifier00816722021427
UDI-Public010081672202142711230518
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K221407
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2025
Initial Report FDA Received Date06/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/18/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability;
Patient SexFemale
Patient RaceWhite
-
-