PERFUSION SYSTEMS HMS PLUS INSTRUMENT; ANALYZER, HEPARIN, AUTOMATED Back to Search Results
Model Number 30514
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to use of the hms plus instruments it was noted that the instruments were failing flag height and receiving abnormal quality control results.The same results were observed when using three different cartridges and instruments.The instruments were used to complete the procedure.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Device evaluation summary: the reported issue that the instrument was failing flag height and receiving abnormal quality control results was verified during service.The service technician noted that the flag height was too high and the dispensing levels were out of limits.The issue was resolved by adjusting the flag height to within limits and adjusting the syringe switch to the proper dispensing levels.Preventive maintenance was performed per specifications.The instrument was analyzed/serviced in the facility by a field service technician.The instrument did not return to a medtronic facility for analysis/service.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic received additional information that both liquids failed.Quality control is performed for this instruments as prescribed by medtronic.There was no error code associated with this issue.No control values obtained.Additional information d4 unique identifier (udi) #: this field was updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic received additional information that the issue was only with two instruments.The second instrument was reported separately and has its own case and service report.The third instrument mentioned underwent preventive maintenance, and no issues were found, therefore, it is not a complaint.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HMS PLUS INSTRUMENT
Common Device Name
ANALYZER, HEPARIN, AUTOMATED
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key22182647
Report Number2184009-2025-00743
Device Sequence Number8617420
Product Code JOX
UDI-Device Identifier00673978698793
UDI-Public(01)00673978698793(11)180910(21)9002714
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K111339
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number30514
Device Catalogue Number30514
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2025
Supplement Date Received by Manufacturer06/12/2025
07/07/2025
Initial Report FDA Received Date06/11/2025
Supplement Report FDA Received Date07/07/2025
07/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/10/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-