| Catalog Number |
KRR100 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the device, krr100, infinity retro-reamer, 10 mm, was being used during an acl repair procedure on (b)(6) 2025 when it was reported, ¿when the surgeon started to drill tibial tunnel retroreamer 10mm broke into the bone.The distal tip of the reamer broke in two parts.Surgeon didn¿t get all the parts and one part was impossible to recover.He did all the tricks to do it but it was not possible.Not sure if the piece is into the patient or removed during the exploration.¿.An alternate same device was used to complete the procedure causing a delay of 20-minutes.There was no report of injury, medical intervention, or hospitalization for the patient.This report is being raised due to the reported injury of possible fragmentation left in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.We will continue to monitor per regulatory requirements to help ensure patient safety.Device not yet received.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned used device, item krr100 was evaluated per print krr100 and found device broken off at the tip.Broken piece was returned for evaluation.The root cause cannot be determined, however, based upon evaluation, photos, and the risk document, the likely cause of this event is likely that the tip was not fully locked in the retro-reaming position during the socket drilling.However, as all parts of the reamer have been returned for the evaluation of the device, the report of fragments being left in the patient is unconfirmed.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 8 complaints, regarding 9 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised not to use excessive force on instruments to avoid damage or breakage during use.Do not use instruments to pry, as bending or breakage may occur.Exercise care in the use of the handpiece holding the reamer to minimize side or bending loads to the reamer which may cause reamer breakage and/or oversized tunnels.Excessive force applied during retrograde drilling can lead to reduced socket sizes.Exercise care in the use of the retro-reamer.Reamers are designed with a sharp cutting edge that could cut the user.We will continue to monitor per regulatory requirements to help ensure patient safety.
|
| |
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the device, krr100, infinity retro-reamer, 10 mm, was being used during an acl repair procedure on 3jun25 when it was reported, ¿when the surgeon started to drill tibial tunnel retroreamer 10mm broke into the bone.The distal tip of the reamer broke in two parts.Surgeon didn¿t get all the parts and one part was impossible to recover.He did all the tricks to do it but it was not possible.Not sure if the piece is into the patient or removed during the exploration.¿ an alternate same device was used to complete the procedure causing a delay of 20-minutes.There was no report of injury, medical intervention, or hospitalization for the patient.This report is being raised due to the reported injury of possible fragmentation left in the patient.
|
| |
|