| Catalog Number |
367326 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D.2b medical device type: one additional code applies; jka e1: initial reporter facility name: (b)(6) a device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using two bd vacutainer® push button blood collection set, the customer noticed damaged in the tubing.There was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received 3 photos for investigation.Evaluation of these photos indicate the reported issue.Additionally, bd conducted visual testing on 100 retained samples for cut and severed tubing.All samples passed with no signs of cut or severed tubing.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: damaged.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that while using two bd vacutainer® push button blood collection set, the customer noticed damaged in the tubing.There was no health impact or consequence reported.
|
| |
|