MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAWIRE DRIVE; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAWIRE DRIVE; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 2077-02
Medical Device Problem Codes Entrapment of Device (1212); Material Deformation (2976); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the rotapro burr became stuck with the rotawire.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified mid right coronary artery.A 1.50mm rotapro and rotawire was selected for use.During the procedure, it was noted the rotawire could not be removed after dynaglide ablation.It was believed that the rotawire was kinked.The rotapro burr became stuck with the rotawire.The rotapro burr was removed outside the patient together with the rotawire.The procedure was completed with another of the same device.No patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device returned for analysis without the burr loaded on it; therefore, no sign of jamming could be found.Body of the guidewire was kinked measured from distal to proximal.Spring tip was observed bent.No other device issues were identified during returned product analysis.
 
Event or Problem Description
It was reported that the rotapro burr became stuck with the rotawire.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified mid right coronary artery.A 1.50mm rotapro and rotawire was selected for use.During the procedure, it was noted the rotawire could not be removed after dynaglide ablation.It was believed that the rotawire was kinked.The rotapro burr became stuck with the rotawire.The rotapro burr was removed outside the patient together with the rotawire.The procedure was completed with another of the same device.No patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE DRIVE
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key22184041
Report Number2124215-2025-35688
Device Sequence Number15625866
Product Code MCX
UDI-Device Identifier08714729996248
UDI-Public(01)08714729996248(17)270203(10)35770715
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2077-02
Device Catalogue Number2077-02
Device Lot Number0035770715
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2025
Supplement Date Received by Manufacturer07/04/2025
Initial Report FDA Received Date06/11/2025
Supplement Report FDA Received Date07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-