| Model Number |
2077-02 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Deformation (2976); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the rotapro burr became stuck with the rotawire.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified mid right coronary artery.A 1.50mm rotapro and rotawire was selected for use.During the procedure, it was noted the rotawire could not be removed after dynaglide ablation.It was believed that the rotawire was kinked.The rotapro burr became stuck with the rotawire.The rotapro burr was removed outside the patient together with the rotawire.The procedure was completed with another of the same device.No patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device returned for analysis without the burr loaded on it; therefore, no sign of jamming could be found.Body of the guidewire was kinked measured from distal to proximal.Spring tip was observed bent.No other device issues were identified during returned product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that the rotapro burr became stuck with the rotawire.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified mid right coronary artery.A 1.50mm rotapro and rotawire was selected for use.During the procedure, it was noted the rotawire could not be removed after dynaglide ablation.It was believed that the rotawire was kinked.The rotapro burr became stuck with the rotawire.The rotapro burr was removed outside the patient together with the rotawire.The procedure was completed with another of the same device.No patient complications reported.
|
| |
|