| Model Number |
WA2UR22A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the tip of the subject device was broken.The issue was identified during set up / inspection for use.There were no reports of patient involvement.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9; g1; g3; h2; h3; h6 and h11.The device was returned to olympus for inspection, and the reported failure was confirmed.In addition, the following reportable malfunction was identified during the device evaluation: distal end came out.A root cause could not be identified.Based on the results of the investigation, the most probable cause was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|