• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 OSSEOTI 3 HOLE SHELL 52MM E; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 OSSEOTI 3 HOLE SHELL 52MM E; PROSTHESIS, HIP Back to Search Results
Catalog Number 110010244
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/19/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: cat# 30103605, lot# 66529256 g7 vit e neutral lnr 36mm e.G2: foreign: australia.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that patient had a primary hip procedure.The patient was implanted with a competitor stem and a zimmer biomet acetabular cup and liner.Subsequently, the patient was experiencing constant pain post-surgery around the buttock area of associated hip replacement.The patient was revised approximately 4 months post-implantation due to pain.All products were removed and replaced.The patient was reimplanted with a competitor stem and zimmer biomet head, cup and liner.It was reported that no further information could be provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: d5; g3; h2; h3; h4; h6 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.It was identified that zimmer biomet devices were implanted with competitor femoral components.Zimmer biomet has not confirmed the compatibility for these combinations of devices.It is unknown if these off-label usages may have caused or contributed to the reported events.Please note, per the ifu, it states "do not use components with components of any other system or manufacturer, unless authorized by zimmer biomet." this complaint cannot be confirmed.A summary of the investigation will be sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 OSSEOTI 3 HOLE SHELL 52MM E
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22185641
Report Number0001825034-2025-01695
Device Sequence Number20974860
Product Code LPH
UDI-Device Identifier00887868355882
UDI-Public(01)00887868355882(17)340715(10)66733530
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K140669
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number110010244
Device Lot Number66733530
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2025
Supplement Date Received by Manufacturer08/06/2025
Initial Report FDA Received Date06/11/2025
Supplement Report FDA Received Date08/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-