• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TRABECULAR METAL REVERSE SHOULDER SYSTEM POLY LINER; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TRABECULAR METAL REVERSE SHOULDER SYSTEM POLY LINER; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number 00434903600
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Discomfort (2330); Bone Notching (4890)
Date of Event 05/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: item#: 115310; lot#: j7292125.G2: foreign - event occurred in australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient underwent a primary shoulder arthroplasty on approximately two (2) years and eight (8) months ago.Subsequently, the patient complained of discomfort and upon assessment, the patient was notching.The patient then underwent a revision surgery approximately one (1) month ago where the surgeon removed a section of proximal gt and bone spurs on the scapula.The surgeon opted to lateralise and increase the offset on the glenosphere to avoid notching and rom restriction/discomfort.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: a reverse-type right shoulder arthroplasty is anatomically aligned.There is no fracture or implant loosening.There is a prominent greater tuberosity and mild prominence of the undersurface of the acromion.There is no evidence of scapular notching.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The complaint is unable to be confirmed.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRABECULAR METAL REVERSE SHOULDER SYSTEM POLY LINER
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22186401
Report Number0001822565-2025-01855
Device Sequence Number12904233
Product Code PHX
UDI-Device Identifier00889024269057
UDI-Public(01)00889024269057(17)290223(10)64812469
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K130661
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number00434903600
Device Lot Number64812469
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2025
Supplement Date Received by Manufacturer06/17/2025
09/17/2025
Initial Report FDA Received Date06/11/2025
Supplement Report FDA Received Date06/18/2025
09/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-