ABBOTT GMBH ALINITY C CARBON DIOXIDE REAGENT KIT; ENZYMATIC, CARBON-DIOXIDE Back to Search Results
Catalog Number 07P72-20
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a1 - patient identifier complete entry = sample id (b)(6) and sample id (b)(6).All available patient information was included.Additional patient details are not available.
 
Event or Problem Description
The customer observed a falsely decreased alinity c carbon dioxide result for one patient with a history of igg k paraprotein.The following data was provided: sample id (b)(6) initial result, on (b)(6) 2025 with serial number (b)(6) and lot number 67438uq12, was 14, result using radiometer blood gas was 28 mmol/l.Sample id (b)(6) initial result, on (b)(6) 2024 with serial number (b)(6) and lot number 66500uq06, was 17, repeat, on radiometer, was 25.4 mmol/l.The customer performed dilution studies which were inconclusive.The customer performed peg studies which showed 167% and the control patient used had 97% recovery.The sample was also tested on the gem and the result was 27.The patient¿s previous results were provided: (b)(6) 2025, at a different site, result was 12 mmol/l; (b)(6) 2024, at a different site, result was 12 mmol/l; (b)(6) 2010 result with a roche assay was 8, repeat, on the gem, was 27.There was no impact to patient management reported.
 
Additional Manufacturer Narrative
The complaint investigation for falsely decreased patient results when using alinity c carbon dioxide, lot numbers 67438uq12 and lot 66500uq06, included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed, as returns were not available.A review of labeling and historical data was done, and both were adequate, with no increase in complaint activity.Manufacturing documentation for the likely cause lots was reviewed and did not identify any issues.Device history record review on lot numbers 67438uq12 and lot 66500uq06 did not identify any non-conformances or deviations with the likely cause lots.Labeling was reviewed and sufficiently addresses the customer's issue.Based on the information provided and abbott diagnostics¿ complaint investigation, no systemic issue or deficiency of alinity c carbon dioxide, lot numbers 67438uq12 and lot 66500uq06, was identified.
 
Event or Problem Description
The customer observed a falsely decreased alinity c carbon dioxide result for one patient with a history of igg k paraprotein.The following data was provided: sample id (b)(6) initial result, on 13may2025 with serial number (b)(6) and lot number 67438uq12, was 14, result using radiometer blood gas was 28 mmol/l.Sample id (b)(6) initial result, on 20nov2024 with serial number ac01718 and lot number 66500uq06, was 17, repeat, on radiometer, was 25.4 mmol/l.The customer performed dilution studies which were inconclusive.The customer performed peg studies which showed 167% and the control patient used had 97% recovery.The sample was also tested on the gem and the result was 27.The patient¿s previous results were provided: 23jan2025, at a different site, result was 12 mmol/l; 02nov2024, at a different site, result was 12 mmol/l; 21jan2010 result with a roche assay was 8, repeat, on the gem, was 27.There was no impact to patient management reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C CARBON DIOXIDE REAGENT KIT
Common Device Name
ENZYMATIC, CARBON-DIOXIDE
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key22187569
Report Number3002809144-2025-00184
Device Sequence Number8691097
Product Code KHS
UDI-Device Identifier00380740121594
UDI-Public(01)00380740121594(17)250731(10)66500UQ06
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K060295
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Catalogue Number07P72-20
Device Lot Number66500UQ06
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2025
Supplement Date Received by Manufacturer06/24/2025
Initial Report FDA Received Date06/11/2025
Supplement Report FDA Received Date06/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY C CO2 RGT 3000T, 07P72-20, 67438UQ12; ALNTY C CO2 RGT 3000T, 07P72-20, 67438UQ12; ALNTY C PROCESSING MODU, 03R67-01, (B)(6); ALNTY C PROCESSING MODU, 03R67-01, (B)(6); ALNTY C PROCESSING MODU, 03R67-01, (B)(6); ALNTY C PROCESSING MODU, 03R67-01, (B)(6)