|
(b)(4).Date sent: 06/11/2025.D4: batch #369d86.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pse45a device was returned with no apparent damage and with no reload present.The reload was received unfired.The device was returned with a non-working firing mechanism.No further functional test could be performed due to the condition of the device.In order to verify the condition of the internal components, the device was disassembled.Upon disassembling, the pcb board was noted to be non-functional.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the pcb board to be damaged.A manufacturing record evaluation was performed for the finished device pse45a lot number c9dw93 and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 369d86, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|