| Catalog Number |
A12MA02501012 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event, a k-file m-access broke during use and fell in mouth.The outcome of the event is not known as of this mdr evaluation.Further information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Summary: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1858178).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique (hand used file - technique no more verifiable to date), overuse (number of uses not communicated), excessive wear, patient condition and behaviour during treatment or material issue (no analysis of the broken fragments possible).In addition, there are further circumstances on the conditions in dentistry (used disinfectants, training/knowledge status), or any other environmental conditions, which are unknown to us and may also have an impact on the reported failure mode.Additional information received that there was no injury that occurred in this event.This is a follow up report for this additional information.This is to correct and remove the codes that were initially reported - removing codes for: health effect - clinical code - 4580.Health effect - impact code - 4648.The correct codes for this complaint are: health effect - clinical code - 4582.Health effect - impact code - 2199.This is a follow up report to correct this code.
|
| |
|
Search Alerts/Recalls
|