|
Reportable based on device analysis completed on 23may2025.It was reported that air leak occurred.The target lesion was located in the left forearm.A 10.0 x40, 75cm gladiator elite balloon catheter was advanced for dilation.During the procedure, an air leakage at the tip of the balloon was noted when the pressure reached at 7 atmospheres.The balloon was replaced with another 10x4, and the procedure was completed after normal dilation.There were no patient complications as a result of this event, and the patient's condition was stable.However, returned device analysis revealed a balloon pinhole.
|