• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ELEKTA SOLUTIONS AB LEKSELL VANTAGE STEREOTACTIC SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ELEKTA SOLUTIONS AB LEKSELL VANTAGE STEREOTACTIC SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Medical Device Problem Code Device Handling Problem (3265)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the vantage frame slipped, noted prior to treatment.
 
Additional Manufacturer Narrative
H4: not known, no information on serial number.A retrospective review of this case based on fda inspection may 22, 2025 has been conducted with the conclusion to file an mdr report for this event.The rigidity of frame fixation is dependent on many different parameters: position of the frame on the head; perpendicularity of the pin versus the skull; the applied force of the pins in to the skull bone; patient anatomy; patient movement; skull bone properties; previous skull bone operations.Complete root cause investigation is pending.
 
Additional Manufacturer Narrative
H11 updated: the root cause was identified to be a use error, in which the firmfix was fixated using too low a torque.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEKSELL VANTAGE STEREOTACTIC SYSTEM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
ELEKTA SOLUTIONS AB
hagaplan 4
stockholm, 113 6 8
SW  113 68
Manufacturer (Section G)
ELEKTA SOLUTIONS AB
400 perimeter center terrace
suite 50
atlanta, ga 30346
SW   30346
Manufacturer Contact
cornerstone
london road
crawley, west sussex RH10 -9BL
MDR Report Key22192869
Report Number3015232217-2025-00030
Device Sequence Number18291124
Product Code HAW
UDI-Device Identifier07340048305167
UDI-Public07340048305167
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K190887
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/12/2025
Supplement Report FDA Received Date07/31/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-