| Catalog Number |
119112 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that when customer performed a pretest before placement on an outpatient, they discovered the foley catheter balloon had been damaged.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed manufacturing related.Received (4) photo samples.The photo sample was returned and could not be evaluated for the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), used silicone foley.Visual inspection of the sample noted using the (in-house) syringe attempt to inflate the catheter balloon with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and the solution went into the drainage lumen and leaked out of the eyelet.Dissected the balloon, double perforation found on perforation notch.This does not meet specification per ip7603444, revision 0 which states "catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.Instructions for use were reviewed for this investigation.The labeling is found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Corrections: d, f, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that when customer performed a pretest before placement on an outpatient, they discovered the foley catheter balloon had been damaged.
|
| |
|