• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH VA LOCKSCR ø2.7 HEAD 2.4 SELF-TAP L58 TA; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH VA LOCKSCR ø2.7 HEAD 2.4 SELF-TAP L58 TA; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 04.211.058S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
J&j medtech is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which j&j medtech has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, j&j medtech or its employees that the report constitutes an admission that the device, j&j medtech, or its employees caused or contributed to the potential event described in additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported the surgeon noted that some metal material shavings came out when he inserted the screws intraoperatively.Surgery date is unknown.It is unknown if there was surgical delay or how the procedure was completed.No known patient consequences.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1 h6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.Device history lot part:04.211.058s lot:386l569 manufacturing site: werk selzach supplier: (b)(4) release to warehouse date:15 apr 2025 expiration date: 01 apr 2035.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If additional information is made available, the investigation will be updated as applicable.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VA LOCKSCR ø2.7 HEAD 2.4 SELF-TAP L58 TA
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach CO
SZ  
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key22193988
Report Number8030965-2025-05878
Device Sequence Number15887708
Product Code HWC
UDI-Device Identifier10886982119216
UDI-Public(01)10886982119216(17)350401(10)386L569
Combination Product (Y/N)N
Initial Reporter CountrySP
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.211.058S
Device Lot Number386L569
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2025
Supplement Date Received by Manufacturer06/16/2025
06/25/2025
Initial Report FDA Received Date06/12/2025
Supplement Report FDA Received Date06/20/2025
06/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORTSCR ø3.5 SELF-TAP L22 TAN.; LWPROF METAPHYS COMPRES SCR ø2.7 SELF-TA.; LWPROF METAPHYS COMPRES SCR ø2.7 SELF-TA.; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L26 TAN.; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L34 TAN.; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L38 TAN.; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L46 TAN.; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L54 TAN.
Patient SexUnknown
-
-