| Model Number |
8300AB |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Code |
Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
05/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 21mm 8300ab intuity valve has been placed under evaluation for a valve in valve procedure after an implant duration of 8 years, 6 months due to severe stenosis.The patient presented with heart failure.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, b5, g3, g6, h2, h6 (health effect: impact code, type of investigation).The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 21 mm 8300ab intuity valve has underwent a valve in valve procedure after an implant duration of 8 years, 7 months due to severe stenosis.The patient presented with heart failure.The tavr was performed with a 23 mm 9750tfx transcatheter valve.The patient was stable post procedure and discharged on pod #1.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, g3, g6, h2, h6 (investigation findings, investigation conclusions).Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including coronary artery disease (cad).H11: corrected data: b2.
|
| |
|