| Catalog Number |
368920 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® fx 5mg 4mg blood collection tubes, an unspecified number of tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary : bd received one photo for investigation, which visually shows that the amount drawn into the tubes is below the fill line on the tube label.A total of 100 retained samples were inspected, and no visible defects were found.Additionally, a functional test was performed on 10 retained samples, and all tubes were found to be within specification limits.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: underfill based on customer photo analysis.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® fx 5mg 4mg blood collection tubes, an unspecified number of tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|