Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
 
Event or Problem Description
It was reported that 4fr single lumen picc fractured.The tear was between the 2cm and 3cm mark with zero being true zero.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak was confirmed.The product returned for evaluation was one 4 fr single lumen powerpicc solo.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue may occur due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated, and a split was observed between the 2cm and the 3cm markings.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: ¿ damage which was circumferentially aligned ¿ fracture edges which were rounded and polished due to repeated material wear ¿ 'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together repetitive kinking of the indwelling catheter appears to have contributed to the observed leak; however, the specific circumstances surrounding how kinking developed into a fracture within the catheter tubing were ultimately unknown.The damage location suggested that catheter securement, access, and maintenance techniques may have contributed.This complaint will be recorded for future trending and monitoring purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC, 3CG
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER (PICC)
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
maddy vincent
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key22196114
Report Number3006260740-2025-03812
Device Sequence Number9189270
Product Code LJS
UDI-Device Identifier00801741034459
UDI-Public(01)00801741034459(11)240415(17)250731(10)REJP4065
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K091324
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 07/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model NumberN/A
Device Catalogue Number1174108D
Device Lot NumberREJP4065
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2025
Supplement Date Received by Manufacturer07/11/2025
Initial Report FDA Received Date06/12/2025
Supplement Report FDA Received Date07/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
MDS-24-5154-C
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-