• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 050-87215A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak on the inside of the cassette door of their cycler after ending their pd treatment.The patient reported receiving an air detected in cassette alarm during drain 2 of 5 of treatment and the treatment was cancelled.Fluid was discovered in the pump module area and on the external of the cassette.The cause of the leak is unknown.The patient was advised to discontinue the use of their cycler for the night to allow it to dry out.Additional information requested but has not been obtained.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak on the inside of the cassette door of their cycler after ending their pd treatment.The patient reported receiving an air detected in cassette alarm during drain 2 of 5 of treatment and the treatment was cancelled.Fluid was discovered in the pump module area and on the external of the cassette.The cause of the leak is unknown.The patient was advised to discontinue the use of their cycler for the night to allow it to dry out.Additional information requested but has not been obtained.
 
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE PATIENT CONN
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key22196565
Report Number0008030665-2025-01339
Device Sequence Number8738115
Product Code FKX
UDI-Device Identifier00840861102709
UDI-Public00840861102709
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K173718
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number050-87215A
Device Lot Number25AR08080
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2025
Supplement Date Received by Manufacturer08/14/2025
Initial Report FDA Received Date06/12/2025
Supplement Report FDA Received Date08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID.; DELFLEX PD FLUID.; LIBERTY SELECT CYCLER.; LIBERTY SELECT CYCLER.
Patient SexUnknown
-
-