• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493892815250
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.A 2.50mm x 15mm maverick 2 balloon catheter was advanced for dilation.However, the balloon burst due to several calcification.The device was removed and completed the procedure with a different device.There was no patient complications reported.Patient was in good condition.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: fg maverick 2 mr, 2.50mm x 15mm was returned for analysis.Visual and tactile inspection revealed no issues identified with the hypotube shaft.A visual inspection of the outer and inner lumen and tactile examination of the entire extrusion found no issues.A detailed microscopic examination of the balloon material identified no tears or pinholes in the balloon.Bumper tip showed no signs of distal tip damage.The device was attached to an encore inflation unit and the balloon was inflated successfully, held pressure and deflated without issue.The encore device was verified before use by using the druck gauge, inflation was to the rate burst pressure (rbp) of 14 atmospheres (atm).
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.A 2.50mm x 15mm maverick 2 balloon catheter was advanced for dilation.However, the balloon burst due to several calcification.The device was removed and completed the procedure with a different device.There was no patient complications reported.Patient was in good condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key22198076
Report Number2124215-2025-38897
Device Sequence Number18469257
Product Code LOX
UDI-Device Identifier08714729370086
UDI-Public(01)08714729370086(17)271226(10)35535911
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH7493892815250
Device Catalogue NumberH7493892815250
Device Lot Number0035535911
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/28/2025
Supplement Date Received by Manufacturer08/26/2025
Initial Report FDA Received Date06/12/2025
Supplement Report FDA Received Date09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexFemale
-
-