| Model Number |
H7493892815250 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.A 2.50mm x 15mm maverick 2 balloon catheter was advanced for dilation.However, the balloon burst due to several calcification.The device was removed and completed the procedure with a different device.There was no patient complications reported.Patient was in good condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: fg maverick 2 mr, 2.50mm x 15mm was returned for analysis.Visual and tactile inspection revealed no issues identified with the hypotube shaft.A visual inspection of the outer and inner lumen and tactile examination of the entire extrusion found no issues.A detailed microscopic examination of the balloon material identified no tears or pinholes in the balloon.Bumper tip showed no signs of distal tip damage.The device was attached to an encore inflation unit and the balloon was inflated successfully, held pressure and deflated without issue.The encore device was verified before use by using the druck gauge, inflation was to the rate burst pressure (rbp) of 14 atmospheres (atm).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.A 2.50mm x 15mm maverick 2 balloon catheter was advanced for dilation.However, the balloon burst due to several calcification.The device was removed and completed the procedure with a different device.There was no patient complications reported.Patient was in good condition.
|
| |
|
Search Alerts/Recalls
|