| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that ps blue impactor pad broke during femoral impaction.The surgical technique was utilized.There was no surgical delay or foreign bodies retained.All available information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6.The event is confirmed.Visual examination of the returned impactor pad identified signs of repeated use (nicked and gouged) and a corner of the impactor pad had fractured off.Not all fractured pieces were returned.The impactor pad fracture surface artifacts of hackle marks and river lines support the overload failure mode and align with a zrm document in which an overload fracture failure mode was identified.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|