Model Number 420006-12
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is in progress to determine the cause an rma was issued to evaluate the intuitive surgical, inc.(isi) device.Of this reported event.Additional information is being gathered to determine the contribution of the device to the customer reported issue.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the pulley is cut during surgery.The procedure was completed with no reported injury.
 
Additional Manufacturer Narrative
An investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The instrument was analyzed and found to have a broken grip cable.When cable breakage occurs, the instrument is immediately inoperable due to loss of functionality.The instrument has tungsten drive cables to transmit motion from the input discs in the housing, through the main shaft, and to the distal instrument tip.These cables are exposed at the instrument tip and control movements at the wrist.The complaint was confirmed by failure analysis, which indicates that the device may have contributed to the customer reported issue.The probable root cause of cable breakage, whether partial or full, is attributed to damage to the cable through external collisions, during use, or during reprocessing.This issue can be resolved by using an alternate instrument to complete the procedure.
 
Event or Problem Description
Refer to h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
LARGE NEEDLE DRIVER
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key22198442
Report Number2955842-2025-25328
Device Sequence Number9739351
Product Code NAY
UDI-Device Identifier00886874111192
UDI-Public(01)00886874111192(10)K10240411
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 05/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420006-12
Device Catalogue Number420006
Device Lot NumberK10240411 0912
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/28/2025
Supplement Date Received by Manufacturer08/26/2025
Initial Report FDA Received Date06/12/2025
Supplement Report FDA Received Date08/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-