|
A non-conformance review was conducted for lot 2415916664 and there were no ncs related to the reported event issued during the manufacturing of this lot.Prior to a lot¿s release, the lot must be deemed acceptable by passing inspections that are based on a valid sampling plan.During production, inspectors routinely examine a statistical sample both physically and visually.There was no sample returned for evaluation, therefore the affected device could not be evaluated to confirm the reported issue.As such, a root cause could not be determined.No corrective action is required at this time.We will continue to monitor related reports to determine if additional actions are necessary.
|