| Catalog Number |
130790030 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the screw tip broke off during implantation.Screw remained implanted.The procedure was successfully completed but it was unknown if there was a surgical delay.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees hat the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: b5.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received: a.What was the screw anchoring? baseplate.B.What was the indication of the procedure? oa.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Investigation summary: screw tip broke off during implantation.Screw remained implanted the product was not returned to j&j medtech orthopaedics; however, photos were provided for review.The photo investigation revealed that dxtend screw lock d4.5x30mm has the body fractured, however with no x-rays provided its not possible to confirm if the fracture fragments remained in the patient.A manufacturing record evaluation was performed for the finished device product description: dxtend screw lock d4.5x30mm product code: 130790030 lot number: 5699309 and no non-conformances or manufacturing irregularities were identified.Although a definitive cause for the implant (screw) fracture is not established, there are many factors that could have contributed to this type of damage, including the possibility of an off-axis assembly or by loosing contact of the screwdriver's tip with the fines of the screw potentially resulting in uneven torque application if protective sleeve is not used, as per delta xtend¿ reverse shoulder system surgical technique care notes (pages 19-23).Therefore, with information provided the mode of failure of the device is multi-factorial and consideration must be given to all other potential influences during surgical process.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the dxtend screw lock d4.5x30mm would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device product description: dxtend screw lock d4.5x30mm product code: 130790030 lot number: 5699309 and no non-conformances or manufacturing irregularities were identified.Device history review: a manufacturing record evaluation was performed for the finished device product description: dxtend screw lock d4.5x30mm product code: 130790030 lot number: 5699309 and no non-conformances or manufacturing irregularities were identified.
|
| |
|