| Catalog Number |
72205058 |
| Medical Device Problem Codes |
Overheating of Device (1437); Optical Obstruction (3002)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during surgery, the camera head heated up and the view became foggy.There was no damage to the patient or user due to the overheating.Surgery was completed with a back-up device, instead.There was no surgical delay, and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h3, h6: the reported device was received for evaluation.A visual inspection revealed moisture inside the lens.A functional evaluation was performed on the returned device and found a foggy video output due to the moisture inside the lens.No overheating was observed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with a component failure.Factors that could have contributed to the foggy video include a failed weld.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|