| Model Number |
8422500 |
| Medical Device Problem Codes |
No Display/Image (1183); Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the screen went blank during use and the user was unsure if the device continued to operate.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was based on the provided log file of the affected device.The reported device behavior could be confirmed.The hardware analysis of previous complaints revealed that the activation of the inrush current limitation of the fuse on the printed circuit board of the lc display led to a dark backlight of the lc display.The activation of this fuse was traced back to adverse tolerances tolerance of the electronic components on the adapter board of the display unit of the device (ecd).In case the backlight of the display unit fails during use the user is no longer able to operate the device.The ventilation performed by the ventilation unit is not affected by a display malfunction and will be continued as set.Safety relevant parameters as fio2, minute volume, or airway pressure are displayed in the secondary oled-display located on the ventilation unit.The inrush current limitation of the fuse has already been increased in order to remedy this issue.The symptom can be resolved by switching off and on the device via the main switch.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|
| |
|