• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL FUSLITEGU_OLYM-ACMI_3.5MMX3.0M; LIGHT, SURGICAL, FIBEROPTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL FUSLITEGU_OLYM-ACMI_3.5MMX3.0M; LIGHT, SURGICAL, FIBEROPTIC Back to Search Results
Catalog Number 242386
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
It was reported from that the fuslitegu_olym-acmi_3.5mmx3.0m device, after the surgery, the connection to the scope had become easily detached.Therefore, the silver connection part remains in the scope side without fail when you detach the device from the scope.This event occurs every time the device is used, causing problems for the surgeon.There was no harm to the patient and no surgical delay.All available information has been disclosed.The exact date of this event was unknown but was noted to have occurred in 2025.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: investigation summary: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that device is in normal use condition, however, on the female end, the connector was found detached.The detached connector was returned.The device was manufactured in march/2023 which is more than 2 years old.A functional test was performed, the female end tip was directed to a light source, the light emission was good and strong.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the fuslitegu_olym-acmi_3.5mmx3.0m would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life; as per the instructions for use; light cables are delicate medical instruments and must be used and handled with care.Process light cables as soon as is reasonably practical following use.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSLITEGU_OLYM-ACMI_3.5MMX3.0M
Common Device Name
LIGHT, SURGICAL, FIBEROPTIC
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key22201380
Report Number1221934-2025-02199
Device Sequence Number7643972
Product Code FST
UDI-Device Identifier10886705028818
UDI-Public(01)10886705028818(10)WO194423
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K901035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number242386
Device Lot NumberWO194423
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2025
Supplement Date Received by Manufacturer06/17/2025
Initial Report FDA Received Date06/13/2025
Supplement Report FDA Received Date07/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/18/2023
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-