This is 2 of 2 reports linked to mfg report number: 3013886523-2025-00142.Event reported form a post-market clinical program: a facility reported a computed tomography (ct) scan revealed a round isodense mass in the pineal region, suggestive of a pineal tumor with significant hydrocephalus.Therefore, on (b)(6) 2021, emergency placement of an external ventricular drain (evd) and intracranial pressure (icp) monitoring were performed under general anesthesia.The patient presented with persistent headache for one week, and on (b)(6) 2021, a hakim valve (id 823832) and a bactiseal catheter kit (id 823072) were implanted.On (b)(6) 2021, ventriculoperitoneal (vp) shunt repositioning surgery was conducted due to suboptimal catheter positioning.The patient's condition stabilized postoperatively, and discharge was completed on (b)(6) 2021.
|
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.Corrected field: g2.The bactiseal catheter was not returned for evaluation as the product is not available for return as per customer; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause(s) of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.Additional information received: the physician has changed the correlation decision of this event from ¿ relevant¿ to ¿ irrelevant¿ (not related to the bactiseal catheter kit - id 823072).
|