|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The perforator was returned for evaluation: dhr - no anomalies were identified that occurred during the manufacture or packaging of the device.Failure analysis - the complaint could not be replicated and passed all visual inspection and functional tests, including failure to disengage and premature disengagement, without any issues.Root cause - a potential cause of the premature disengagement may be related to the angle positioning, pressure and/or incorrect fit between the hudson driver and the perforator body application during use.
|