| Model Number |
HIP LABRAL SCORPION |
| Medical Device Problem Codes |
Positioning Failure (1158); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 05/19/2025, it was reported by a sales representative via (b)(4) that an ar-16991 hip labral scorpion was being utilized for an arthroscopic hip labral repair utilizing knotless fibertak anchors.The suture fell out of the top jaw 3 times, misfired twice, and cut the repair stitch upon passing twice.The device was changed out with a new device (keeping the same needle) and were able to put in 3 subsequent anchors without issue.There is an issue with the top jaw of the scorpion causing it to limit its function.This was discovered during the case with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to wear and tear incurred from repeated/extended usage in the field.Manufactured date: 15-feb-2016.
|
| |
|