| Catalog Number |
STP-AT-012 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Hypoglycemia (1912)
|
| Date of Event |
05/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter occurred.The sensor was inserted into the arm on (b)(6) 2025.Data was evaluated and the allegation was undetermined.The probable cause could not be determined via data.The reported glucose values fall within the d zone of the parkes error grid.The data investigation did not find blood glucose calibration values to compare to cgm values during the reported sensor session or the calibrations during the reported sensor session were in accurate range per device specifications.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Search Alerts/Recalls
|