| Catalog Number |
VCP603H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 6/13/2025.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, the following was obtained: how many devices demonstrated the alleged deficiency in each surgery? 1 device per surgery.When were the needles detached/pulled off from the sutures (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify the needles detached off from the suture during use on the patient.Could you kindly perform the follow-up attempt for the product return? please ensure that the shipment tracking number is provided.As soon as i receive the return kit, i¿ll make sure to provide you with the tracking number.Please provide the source or name of person providing answers to follow-up questions: you can follow up with me and i¿ll follow up with the person responsible.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The single complaint was reported with multiple events.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a general surgery on (b)(6) 2025 and suture was used.During the surgery, the suture detached easily from the needle.This happened twice this week with this lot number, and they could have lost the needle in the patient's abdomen.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device 10155x / vcp603h65 batch number, and non-conformances no related to the reported complaint condition were identified.
|
| |
|
Search Alerts/Recalls
|
|
|