D6a: implanted date: device was not implanted.D6b: explanted date: device was not explanted.The actual sample was not available for evaluation.Instead, a reference video was received.The sg2 syringe was observed to be activated by pressing the sheath on a hard surface after multiple attempts at activation.Retention samples were visually inspected and found to be free of any damage or irregularities on the sheath and collar.The sheath activation test was performed on the samples, and all passed.There was no issued nonconformity report related to human error during lot production.There are no nonconformities in the assembled needle lot supplied during production.During the collar assembly process, the hub fit is loosened from the protector before placing the needle into the collar.Before proceeding to the next process, the hub, cannula, and collar are pressed into the protector, wherein a photoelectric sensor detects if the height of the protector is correct.Our safety needle features a hinged safety sheath attached to the needle hub.The safety sheath contains a locking mechanism that is activated when manually pressed over the needle immediately after use and just before disposal to minimize the possibility of sharps injury.The sg2 safety sheath is activated with a one-handed operation by pressing the sheath on a hard surface with firm and quick motion.An audible and/or tactile "click" indicates activation, which can also be visually confirmed.The collar undergoes an initial product inspection check (ipic) before mass production, during raw material change, mold maintenance, plant shutdown, cavity closure, and mold/product type change.Any molding defects that may affect product function are part of the inspection item.We also have a visual in-process and product confirmation inspection to ensure the collar is in good condition.Visual in-process inspections are conducted during the manufacturing process to detect abnormalities in the product that may cause issues during sheath activation.The critical locking part of the sg2 product is checked for defects such as short shot, sheath collar fitting, and over or under insertion of the collar into the hub.We have functional tests conducted to confirm that our products are free of defects that would lead to difficulty or failure of device activation.We perform safety sheath activation during sg assembly inspection and outgoing inspection of finished products, where the cannula should be captured under the sheath's tooth.During the component fit test, we also check that the sheath lugs are securely attached to the collar ear.Before shipment, qc conducts outgoing inspections to check the overall condition of the products.The product's instructions for use (ifu) provide detailed instructions for using the surguard2 hypodermic syringe with safety needle device, including warnings and precautions.The collar is manufactured in-house with validated tpc molding machines.There were no nonconformance reports issued for the supplied molded part lots.Our safety sheath has undergone incoming inspection which includes visual inspection, dimensional inspection, and verification of certificate of conformity.From the previous two fiscal years to the present, we have received a total of five (5) complaints for the specific item code.The cause of these complaints was not identified as being related to our product or the manufacturing process.Retention samples of 23g were simulated.The samples were activated by pressing the sheath on a hard surface with a firm and quick motion.All the samples were effectively activated on one attempt, and the cannula was captured under the sheath tooth.There were no irregularities or bending of the cannula encountered during the activation.Based on the results of our investigation, the root cause of the complaint is not related to our product or manufacturing process.The video provided showed that the sg2 syringe was activated after multiple attempts.However, when retention samples were evaluated, all the samples were successfully activated on one attempt, confirming that there were no issues with the sheath's locking mechanism.The retention samples were inspected and confirmed free from defects that will affect the activation of the safety sheath.Also, the lot history file showed no non-conformity or any related troubles that will affect the product quality.The corrective action was not applicable.The root cause of the complaint could be a user error.The reported device was activated after multiple attempts, which is unlikely to happen if the sg2 syringe was activated by pressing the sheath on a hard surface with firm and quick motion.A review of the product's history file revealed no related issues that would cause sheath activation failure.We perform routine inspections to confirm the performance of the device throughout its manufacturing process; this includes visual and functional checks.The components that are critical to the product's safety activation are checked to ensure that no defects occur that could result in sheath activation issues.We recommend following the instructions for use (ifu) for proper use of the sg2 syringe with safety needle, which includes caution, precautions, and warnings to avoid problems during sheath activation.Terumo medical corporation (tmc) (importer) registration no.(b)(4) is submitting this report on behalf of terumo (philippines) corporation (manufacturer) registration no.(b)(4).
|