| Catalog Number |
151820041 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
04/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a septic mobilization of right knee.No other information was provided at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees hat the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d10.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Product description- attune medial dome pat 41mm.Product code-151820041.Lot no- 4634595.Quantity of manufactured- (b)(4).Date of manufacturing-06-dec-2024.Expiry date-30 nov 2029.Any anomalies or deviations identified in dhr: no.Ifu #: 090200828 ifu attune fixed bearing.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: product description- attune medial dome pat 41mm.Product code-151820041.Lot no- 4634595.Quantity of manufactured- (b)(4).Date of manufacturing-06-dec-2024.Expiry date-30 nov 2029.Any anomalies or deviations identified in dhr:no.Ifu #: 090200828 ifu attune fixed bearing.Device history review: a manufacturing record evaluation was performed for the finished device [lot/serial/batch] number, and no non-conformances / manufacturing irregularities were identified.Corrected: h4.
|
| |
|
Search Alerts/Recalls
|