|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3542557MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Local Reaction (2035)
|
| Date of Event |
04/15/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
On 29-aug-2025, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced a rupture in the breast implant.In addition, the patient experienced swelling and pain.The product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned device revealed that the breast implant had an area of siltex cracking on the edges of the tear, measuring 4.8 cm, on the posterior view.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.Swelling is a temporary normal post-operative condition.Depending on presentation, delayed swelling may require additional work-up by the surgeon.Swelling is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis rupture, breast swelling and pain.D6b explantation date: (b)(6) 2025.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a primary breast augmentation procedure with a mentor memorygel breast implant 255cc gel breast prosthesis that ruptured post implantation.Rupture of the patient¿s right breast prosthesis was reported.Additionally, the patient developed swelling and pain in the right breast.As a result, the patient is scheduled to undergo explantation on (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
On 14-jul-2025, mentor received additional information about the event.The patient had both breast prostheses explanted on (b)(6) 2025 and they were replaced with mentor memorygel xtra breast implant 405cc gel breast prostheses.On 30-jul-2025, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|