AESCULAP AG AS VEGA SYSTEM® PS; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMICONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number NX036Z
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/13/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The adverse event is filed under aesculap reference number: (b)(4).Investigation results: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.Batch history review: the device history records (dhr) were not able to be reviewed as no lot number was made available.However, all device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.Conclusion and measures/preventative measures: a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Additional Manufacturer Narrative
The adverse event is filed under aesculap reference no.(b)(4).Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that there was an issue with a component of columbus knee system.According to the complaint description, this knee implant was "defective and failed prematurely because the ceramic coated surface fails to properly adhere to the cement".There was postoperative mechanical loosening.It was noted that the surgeon easily removed the implant by hand during the revision.Additional information was not provided.
 
Additional Manufacturer Narrative
The adverse event is filed under ais reference no.Xc 100039363 cc 400712790 updated information: b3 date of event b5 event description d1,d2,d4 suspected medical device information d6a d6b implant and explant dates investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
The patient underwent revision total knee arthroplasty of the right knee for loosening and failure of prior components.The femoral component was removed with micro-osteotome techniques.The tibia demonstrated medial-sided collapse and was reamed/broached to appropriate size.Trial components were placed, and stable fixation with acceptable flexion/extension was achieved.Final femoral and tibial components were implanted using third-generation cement technique.A polyethylene insert was placed, and the joint was copiously irrigated.The patient tolerated the procedure, was awakened from anesthesia, and transferred to pacu in stable condition.
 
Additional Manufacturer Narrative
The adverse event is filed under ais reference no.(b)(4).Additional information: b7 patient medical history.E1 initial reporter updated.
 
Additional Manufacturer Narrative
The adverse event is filed under ais reference no.(b)(4).Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.The device history records (dhr) were reviewed for the available lot number; the device was found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Additional Manufacturer Narrative
The adverse event is filed under ais reference no.(b)(4).Investigation results: the complained medical device was not made available for investigation.The investigation was based upon historical data analysis, event description and review of a pictures.The provided pictures show the following situation: the femoral component (nx036z) on the right-side shows bone cement adhesive on nearly the whole intended area.The tibial component (nx057z) on the right-side shows nearly no bone cement residue.The tibial- as well as the femoral component (nx058z and nx016z) on the left side show nearly no bone cement residue.The device quality and manufacturing history records have been checked for the available lot number and were found to be according to the manufacturer's specifications, valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against the affected batch number.Conclusion/preventive measures: on the basis of the current information and without the complained medical device being available for investigation, a clear root cause cannot be drawn.In the event that the complained medical device will be provided for investigation in the future, an update of this report will be provided unsolicited.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA SYSTEM® PS
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMICONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key22205756
Report Number3005673311-2025-00043
Device Sequence Number20524278
Product Code JWH
UDI-Device Identifier04046963810245
UDI-Public(01)04046963810245(11)160225(17)260131(10)52219604
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 04/06/2026
Report Date (Section F) 06/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model NumberNX036Z
Device Catalogue NumberNX036Z
Device Lot Number52219604
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date05/22/2025
Event Location Hospital
Date Report to Manufacturer05/22/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Not provided
Not provided
Not provided
Initial Report FDA Received Date06/13/2025
Supplement Report FDA Received Date06/27/2025
08/27/2025
09/30/2025
12/09/2025
04/06/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexUnknown
Date Report Sent to FDA06/13/2025
-
-