BOLTON MEDICAL, INC RELAY PRO NBS THORACIC STENT-GRAFT SYSTEM; STENT, ENDOVASCULAR GRAFT, AORTIC
|
Back to Search Results |
|
| Catalog Number |
28-N4-30-104-34U |
| Medical Device Problem Codes |
Difficult to Advance (2920); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
05/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
"upon insertion of the relay pro delivery system it was noted that the sheath tip did not have a smooth transition to the nose cone.The physician attempted to fix the transition by rolling and pressing the sheath down around the nose cone.Upon insertion the sheath caught the cfa and tore the vessel.The device was removed and balloon control was achieved.A cutdown was performed and the vessel was repaired surgically.Patient is stable with no further issues." patient outcome: "cfa tear treated with surgical cutdown, viabahn stents, and surgical ptfe graft.".
|
| |
|
Additional Manufacturer Narrative
|
|
Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Additional Manufacturer Narrative
|
|
Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
"upon insertion of the relay pro delivery system it was noted that the sheath tip did not have a smooth transition to the nose cone.The physician attempted to fix the transition by rolling and pressing the sheath down around the nose cone.Upon insertion the sheath caught the cfa and tore the vessel.The device was removed and balloon control was achieved.A cutdown was performed and the vessel was repaired surgically.Patient is stable with no further issues." patient outcome: "cfa tear treated with surgical cutdown, viabahn stents, and surgical ptfe graft.".
|
| |
|