MAUDE Adverse Event Report: EDWARDS LIFESCIENCES KONECT RESILIA AORTIC VALVED CONDUIT AVC; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES KONECT RESILIA AORTIC VALVED CONDUIT AVC; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 11060A
Medical Device Problem Codes Insufficient Device Problem Information (3190); Device Stenosis (4066)
Health Effect - Clinical Codes Chest Pain (1776); Dyspnea (1816); Insufficient Information (4580)
Date of Event 05/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned through implant patient registry that a patient with a 22mm 11060a aortic valve underwent a valve-in-valve procedure after an implant duration of 2 years, 8 months due to unknown reason.The procedure was performed with a 26mm 9755rsl transcatheter valve.The patient was in stable condition post procedure.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional customer complaint.The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
Updated sections: b5, b7, g3, g6, h6 component code, health effect - clinical code, device code(s), type of investigation, investigation findings, and investigation conclusions.The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including hyperlipidemia, coronary artery disease.
 
Event or Problem Description
It was learned through implant patient registry and investigation that a patient with a 22mm 11060a aortic valve underwent a valve-in-valve procedure after an implant duration of 2 years, 8 months due to severe stenosis.The patient presented with chest tightness and shortness of breath.The procedure was performed successfully with a 26mm 9755rsl transcatheter valve.The patient was discharge on pod#3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KONECT RESILIA AORTIC VALVED CONDUIT AVC
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key22207560
Report Number2015691-2025-04874
Device Sequence Number6755929
Product Code LWR
UDI-Device Identifier00690103201390
UDI-Public(01)00690103201390(17)240412
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P150048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/12/2024
Device Model Number11060A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2025
Supplement Date Received by Manufacturer06/30/2025
Initial Report FDA Received Date06/13/2025
Supplement Report FDA Received Date07/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age55 YR
Patient SexMale
-
-