Model Number PM2110
Medical Device Problem Codes Premature Discharge of Battery (1057); Failure to Interrogate (1332); Inappropriate or Unexpected Reset (2959)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented in-clinic for a routine check-up.Upon interrogation it was noted that the pacemaker exhibited premature battery depletion, and the right-atrial (ra) lead and the right-ventricular (rv) both exhibited loss of capture and low pacing impedance.Programming changes were performed, but both leads pacing impedances remained unchanged.Then, the pacemaker spontaneously went into backup operation and failed to be interrogated.No further intervention was performed.The patient condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCENT DR
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key22207801
Report Number2017865-2025-85895
Device Sequence Number17831489
Product Code NVZ
UDI-Device Identifier05414734503600
UDI-Public(01)05414734503600(10)4164425(17)150131
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2015
Device Model NumberPM2110
Device Lot Number4164425
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2025
Supplement Date Received by Manufacturer09/23/2025
Initial Report FDA Received Date06/13/2025
Supplement Report FDA Received Date09/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/16/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age87 YR
Patient SexMale
-
-