| Catalog Number |
762165 |
| Medical Device Problem Code |
Coagulation in Device or Device Ingredient (1096)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D4.Medical device expiration date: unknown.H4.Device manufacture date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported after freezing paxgene® blood rna tube, clotting was seen at the bottom of an unspecified number of tubes.No adverse impact was reported regarding the patient or user.
|
| |
|
Event or Problem Description
|
|
It was reported after freezing paxgene® blood rna tube, clotting was seen at the bottom of an unspecified number of tubes.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary : bd had not received any samples or photographs for investigation.Complaints for sample quality are under statistical control for the month of may 2025.At this time, further testing is not indicated.The device history records could not be reviewed as the lot number is unknown.This complaint has not been confirmed for the indicated failure mode: clotting.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|