No technical inquiries were received from the customer.One opened probe was received, with a tip protector, in a tray, for the report.The sample was visually inspected and found to be conforming.The sample was then functionally tested for actuation and cut.The sample was found to be conforming for actuation and nonconforming for cut.The probe was disassembled and the components inspected.Nominal wear observed on inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Wear marks observed at bend area and cutting edge of the inner cutter and gouge marks observed at other area on the inner cutter.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification (rfid) tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.The complaint evaluation confirms that the probe had a cut failure.The most likely cause for the poor cutting is the observed gouge marks on the cutting edge of the inner cutter of the probe.A damaged cutting edge can decrease the quality of the cut performed by the probe.How and when the cutting edge of the inner cutter of the probe became damaged cannot be determined form this evaluation.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time as the exact root cause of the cut failure could not be determined.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: 2025-40488.
|