| Catalog Number |
119216M |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon on four different foleys catheters deflated and fell out of the patient.(lot#ngjq1070 and lot#ngjr1840).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.Instructions for use were reviewed for this investigation.The labeling is found to be adequate as it states that: "proceed with catheterization in usual manner using the dominant hand.When catheter tip has entered bladder, urine will be visible in the drainage tube insert catheter two more inches and inflate catheter balloon.Inflate catheter balloon using entire 10cc of sterile water provided in the prefilled syringe note: use of less than 10cc can result in asymmetrically inflated balloon once inflated, gently pull catheter until the inflated balloon is snug against the bladder neck secure the foley catheter to the patient." a review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Corrections: d, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon on four different foleys catheters deflated and fell out of the patient.(lot #ngjq1070 and lot #ngjr1840-row).Per customer via email on 26jun2025, tm for the hospital, they advised that they had returned the product for evaluation.As of now no additional information was available.
|
| |
|
Search Alerts/Recalls
|