| Model Number |
H938724E |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an exactamix 2400 valve set was leaking from port 1.The leak contributed to the prepared bags being contaminated with the components at port 1 (water soluble vitamins).The valve strip was replaced to resolve the issue.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h4 and h6.H4: device manufacture date - the lot was manufactured between may 24-27, 2024.H11: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|