|
|
| Catalog Number |
Z311H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) h6 component code: g07002 - device evaluation pending.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The device upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.Additional information was requested, and the following was obtained via: when did the 13 needles detach/pull off from the suture (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please explain.Unk please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.The device has been received at (b)(6) and will be shipped.Please check rmao.Tracking number: (b)(4).Provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq).(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown surgery on an unknown date and suture was used.During the procedure, it was reported by a sales rep that, during an unknown surgery, the needle was detached from the suture easily.It was occurred with 13 packages.It was not a control release needle.Another product was used to complete the case.When we send the sample to you, we will let you know its return date and tracking number.Further details are not provided from the hp.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 7/11/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h3, h6.H3 investigational summary: the product was returned to ethicon for evaluation.Five representative unopened samples were received for analysis.Product code sxpp1b406.In order to evaluate the condition of the returned samples, the packets were examined for visual defects: appearance, color package, wrinkles in the seal area, continuous seals, seal margins, over-sealing, damage (torn, punctured) and none was observed.The packets were opened, and the swage and attachment area were noted to be as expected.During the visual inspection, the suture was correctly placed on the winding former.Suture was dispensed without problems and examined along the strand and no anomalies could be observed.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|