| Model Number |
Z-800F |
| Medical Device Problem Code |
Pressure Problem (3012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Intuvie received a report indicating that an infusion pump failed the 30psi occlusion test, recording a pressure of 12psi, which is outside the acceptable range of 22 to 38 psi.No patient involvement was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuvie received a report indicating that an infusion pump failed the 30psi occlusion test, recording a pressure of 16psi which is outside the acceptable range of 22 to 38 psi.A review of the device's serial number history revealed no similar complaints.The failure mode was confirmed, and the probable cause was determined be an out-of-calibration condition.The issue was successfully resolved by recalibrating.Capa-zm2023-23 was initiated to investigate the root cause for this failure mode.The occlusion failure was identified during servicing performed by a third-party service provider.Reference to complaint# (b)(4).
|
| |
|