• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN MULTIHOLE W/GRIPTION 60MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN MULTIHOLE W/GRIPTION 60MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121730060
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870)
Date of Event 05/22/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.No device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process; all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.
 
Event or Problem Description
It was reported that the patient had surgery on (b)(6) 2025.Patient fell and pushed new acetabular through medial wall into their pelvis.Acetabular component was retrieved including two screws and poly still locked into the cup.New dpx revision cup was implanted with five screws.New 40mm poly and ceramic ts revision head were implanted.Doi: (b)(6) 2025.Dor: (b)(6) 2025.Affected side: left hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MULTIHOLE W/GRIPTION 60MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22214340
Report Number1818910-2025-09806
Device Sequence Number18307596
Product Code KWA
UDI-Device Identifier10603295010029
UDI-Public(01)10603295010029(17)20290630(10)J4319N
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K093646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number121730060
Device Lot NumberJ4319N
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/22/2025
Initial Report FDA Received Date06/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE HD CER 40 +5.; PINN CAN BONE SCREW 6.5MMX30MM.; PINN CAN BONE SCREW 6.5MMX30MM.; UNKNOWN HIP ACETABULAR LINERS.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-