US SURGICAL PUERTO RICO EEA; STAPLER, SURGICAL Back to Search Results
Model Number EEA31
Medical Device Problem Codes Difficult to Remove (1528); Failure to Form Staple (2579)
Health Effect - Clinical Code Failure to Anastomose (1028)
Date of Event 05/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a laparoscopic sigmoid colectomy, when using the new circular stapler to anastomose the descending colon and rectum, poor staple formation was observed and the suture was subsequently used to reinforce stitching.It was noted that the anvil did not tilt after firing.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the green bar was not visible in the indicator window and the safety feature was not engaged.Inspection under the microscope displayed divots on the knife blade.The anvil of the instrument was tilted and attached to the instrument.Further inspection of the anvil cutting ring showed shallow traces of knife impression and the ring was received partially severed.It was reported that poor staple formation was observed.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the instrument handle compression is not completed or if the instrument is applied against an excessive amount of tissue during the clinical application.It was also reported that the anvil did not tilt after firing.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: when firing the stapler, make sure to fully squeeze the instrument handle until the metal underside of the handle contacts the stapler body to the fullest extent.Partial or incomplete handle squeezes may result in unacceptable staple formation and/or incomplete knife cut.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEA
Common Device Name
STAPLER, SURGICAL
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key22214637
Report Number2647580-2025-01835
Device Sequence Number15640604
Product Code GAG
UDI-Device Identifier10884523005554
UDI-Public10884523005554
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K221003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEEA31
Device Catalogue NumberEEA31
Device Lot NumberP3G0130S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2025
Supplement Date Received by Manufacturer07/29/2025
Initial Report FDA Received Date06/16/2025
Supplement Report FDA Received Date08/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
Patient Weight68 KG
-
-